In the electronics and electrical industry, there are requirements for controlling harmful organic volatile compounds present in materials such as plastics The Apollo series is manufactured to test DIBP DBP BBP DEHP for RoHS 2 0 Compliance

The principle

The instrument consists of three core components: a pyrolyzer, a gas chromatograph, and a data processing system.

The weighed sample to be analyzed is fed into the pyrolysis chamber, where it is rapidly heated and desorbed under a protective gas atmosphere. This causes the sample to quickly transform from a solid to a liquid state, and all its phthalate components are instantly vaporized. These vaporized components are then carried by the carrier gas into the chromatographic column. After being separated by the chromatographic column, they enter the detector (FID) one after another for detection. The detection signals of each component are processed using algorithms to calculate their respective contents in the sample.


Specification

Pyrolysis chamber

     Power: 15VDC; 200W

      Temperature control range: 100℃ - 600℃, temperature control accuracy: ±1℃; heating rate: 10℃/s - 100℃/s, set gradient: 5℃/s

Chromatography

lChromatographic Column Oven

         Temperature control range: 50℃ - 350℃, temperature control accuracy: ±1℃; heating rate: 10℃/min - 30℃/min

         the temperature gradient: 5℃/min; Cooling rate: from 350℃ to 50℃ within 10 minutes;

lColumn

        30m special capillary column for phthalates

lDetector

         Self - made high - sensitivity FID detector, detection limit: 5×10⁻¹² g/s (n - hexadecane)

Operation Software

lFully digitalized operation. It has a universal analysis condition module; it is also possible to manually program the temperature increase procedures for pyrolysis and chromatographic separation, and compile and save personalized analysis condition modules.

lReal-time monitoring and automatic online data collection, and automatic saving of operation logs

lAutonomously identify the target components, or manual adjustment can also be carried out.  

lAutomatically perform data processing, and manual adjustment of the data processing method is also possible.

lAutomatically generate templated result reports, and manual adjustments can also be made.

lWith manual spectrogram comparison function

lHierarchical software permission management

Sampling

     Manual + Automatic (equipped with a 22-position automatic sampler)